Permeability Teaster

Porous Materials Inc Products - Permeabilitiy Tester


Gas Permeability Tester
Permeability of gases, vapors, and liquids is important for many industrial applications. Permeametry involves measurement of flow rate of a fluid across a sample and the pressure drop across it in order to compute permeability using Darcy’s law. Gas flow rates and pressure drops are also used to compute many other useful characteristics of porous media.
The PMI Gas Permeameter provides fast and accurate measurement of gas permeability of solid/hollow cylindrical and sheet samples. Featuring nondestructive testing and fast results, PMI’s Gas Permeameter is perfect for both R&D and quality control.

Feature

  1. Windows-based software handles all control, measurement, data collection, and report generation; complete manual control also possible.
  2. Compatible with Windows 95 or higher
  3. Real-time graphical test display depicts testing status and results throughout operation
  4. Non-destructive testing
  5. Length of test approximately 10 minutes
  6. Wide range of acceptable sample types and sizes
  7. Multiple sample chambers available
  8. Minimal maintenance required
  9. Units in Darcy, Frazier, Gurley, and many others

Principle

    Gas permeability is defined by Darcy’s law. According to this law, flow of fluids through porous media is proportional to the pressure gradient causing flow. The proportionality constant normally decreases with increase in viscosity of the fluid. Therefore, incorporating viscosity in the equation Darcy’s law becomes:

Function

  • Industries worldwide utilize the PMI Gas Permeameter for R&D and quality control.
  • Applicable industries include: Automotive, Battery, Biotechnology Separator ,Ceramic, Filtration, Fuel Cells, Geotextiles, Nonwovens, Paper, and Textiles.

Sample Size
  • Dia. 5cm2~100cm2 (Others avalible)
Permeability Range
  • 1 x 10-3 - 50 darcies
Clamping Design
  • Manual type or Pneumatic type (180N)
Test Gas
  • N2, Air or others
Pressure Range
  • 0~500 psi (Others avalible)
Flow Range
  • 10cc/min ~ 200 L/min (Others avalible)

Data